Before the patient · 4 min

Preflight that actually matters

The HAMILTON-T1® is a turbine ICU ventilator that happens to fit in a helicopter. Treat the preflight like you would a flight control check — not a vibes check.

Power and gas

Confirm AC or inverter, battery state, and that the turbine spins without a wall air hose. The T1 does not need piped air. It does need a working O2 source and a planned bottle duration.

Circuit and leak

Adult or peds circuit, HME or heated circuit per program, expiratory valve seated. Run the tightness test. A failed test on the pad is cheaper than a disconnect alarm at 3,000 ft.

Patient profile first

Sex and height write the IBW that every mL/kg number depends on. Leaving the last patient's 175 cm male in the T1 is how a 160 cm woman gets 9 mL/kg all night.

Alarms before you leave the bay

Pmax / Plimit, low VTE, low MV, apnea time. If you inherit ICU limits of VTE 80 mL you will never hear the disconnect.

Flight Crew Buddy LLC

Training the best to take care of the rest

Flight Crew Buddy trains crews who use the HAMILTON-T1®

Independent product of Flight Crew Buddy LLC · flightcrewbuddy.com

HAMILTON-T1®, ASV®, and Hamilton Medical® are trademarks of Hamilton Medical AG and/or its affiliates.

Flight Crew Buddy LLC is not affiliated with, endorsed by, sponsored by, or a training partner of Hamilton Medical AG. This is an independent training aid. Not for clinical use. Not a substitute for manufacturer training or clinical judgment.

The on-screen ventilator is a training layout — not a photograph of the HAMILTON-T1® and not Hamilton software. We will not chase a photo of the box. Crews set the breath from IBW, the lung, sedation, and the gas. ASV® is here so you can recognize it, not as the Flight Crew Buddy default.

© 2026 Flight Crew Buddy LLC. All rights reserved.